What is the minimum applied voltage necessary to maintain either full current or full dynamic load settings?

The 63600 Loads Minimum Applied Voltage across the Load Terminals required to maintain Full Current load settings are:

  1. 63610-80-20  Minimum applied Voltage of 0.5V to maintain 20A Load
  2. 63630-80-60  Minimum applied Voltage of 0.5V to maintain 60A Load
  3. 63640-80-80  Minimum applied Voltage of 0.4V to maintain 80A Load
  4. 63630-600-15 Minimum applied Voltage of 2.0V to maintain 15A Load

The Load will continue to conduct current to a much lower voltage with some de-rating of current see chart below:

De-rating graph for full current operation.


The 63600 Loads Minimum Applied Voltage across the Load Terminal required to maintain Full Dynamic load settings are:

  1. 63610-80-20 Minimum applied Voltage of 1.5V to maintain 80 Amp Pulse Current at 20KHz
  2. 63630-80-60 Minimum applied Voltage of 1.5V to maintain 60A Load Pulse Current at 20KHz
  3. 63640-80-80 Minimum applied Voltage of 1.5V to maintain 80A Load Pulse Current at 20KHz
  4. 63630-600-15 Minimum applied Voltage of 3.0V to maintain 15A Load Pulse Current at 20KHz

The Load will continue to conduct current at a lower voltage with a reduced current setting (Load 1 & Load 2) and increased Time (T1 and T2) see chart below:

dynamiccurrent


For more information, refer to the datasheet or user manual.