Ground Bond Tests

The purpose of a ground continuity test is to verify that all conductive parts of a product that are exposed to user contact are connected to the power line ground (the “green” wire).

The theory is that if an insulation failure occurs that connects power line voltage to an exposed part and a user then comes into contact with that part, current will flow through the low resistance ground path to the green wire, tripping a circuit breaker or blowing a fuse, rather than flowing through the higher resistance of the user’s body. Connecting all exposed conductive parts solidly to ground safely diverts the current away from the person.
19305 Impulse Winding Tester

Impulse Winding Tester

Chroma 19305

10bit / 5ns (200MHz)

More than 10uH

LAN, USB, and RS232

Specifically designed to test winding components using a high-voltage charged micro capacitor and the coil under test, forming an RLC parallel resonant circuit.

Partial Discharge Tester – 19501-K

Chroma 19501-K

Partial Discharge Test
1-200 pC

0.10 kV~10.00 kV, steps 0.01 kV

LAN, USB, and RS232

Specifically designed for testing high voltage semiconductor components and high insulation materials.

Guardian Ground Bond Tester

Chroma 19572

AC: 3~45A

0.1~510 mΩ

50Hz / 60Hz

Provides high current testing of ground continuity between chassis and power cord ground on a wide variety of electrical products and appliances

Electrical Safety Analyzer

Chroma 19032

AC: 0.05~5kV
DC: 0.05~6kV

DC: 12mA , AC: 40mA (19032)
DC: 25mA , AC: 100mA (19032-P)

0.1MΩ~50GΩ / 1kV max

The test-to-any-standard electrical safety analyzer. Easy, precise testing. A must for medical device testing and IQOQ compliance.

Polarization Test

A polarization test is usually performed as part of one of the other tests, such as a line voltage leakage or a hipot test. It is a simple test that verifies that a product supplied with a polarized line cord (either a 3-prong plug or a 2-prong plug with the neutral prong larger than the other) is properly connected.

The test may be just a visual inspection or it may be a wiring continuity check. A main purpose of such a test is to ensure that the line and neutral conductors are not interchanged.

Ground Continuity Test

Since many older homes may be wired as 2-wire systems without solid ground connections, regulatory agencies require all products manufactured with 3-wire cords to pass the same hipot tests as ungrounded products. In such cases, the user is protected by the electrical insulation rather than by the safety ground.

Ground continuity tests are normally performed with a low current DC signal that checks to ensure that the ground connection has a resistance of less than 1 ohm. Ground continuity testing is not only helpful in determining how well a product will fare during a laboratory investigation, but also is useful in a production line environment to ensure quality and user safety.

Ground Bond Test

Ground bond testing requires application of a high current source to a conductive surface of the product and measurement of the voltage drop across the ground connection. This is to determine that bonding is adequate and that the circuit can carry the specified current safely. One common method of ground bond testing, applies a 25A source between the protective grounding terminal of the device and all conductive parts that are accessible to the user. The tester used for this purpose supplies the required current and displays the ground circuit resistance in ohms or milliohms.

Because the resistance to ground is usually a very low value, the resistance of the connecting leads from the tester itself can cause errors in the measurement. Such errors can be corrected either by measuring the resistance of the leads before the test and then subtracting that value from the test value or by using a “Kelvin” test setup. A Kelvin connection automatically compensates for the lead resistance by bringing an extra lead to the point of measurement. The extra lead is connected so as to balance out the resistance of the test lead. Most standards recommend a ground resistance of <100 milliohms, excluding the power cable.