Question: Why is Accuracy and Multiple Ranges important?

Answer:  Most accuracy specifications for Test Equipment are listed as % rdg + % fs.

Example: 0.1% rdg + 0.1% fs

This is a percentage of the reading or measurement + the percentage of the Full Scale for the range you are in.

If a DC Electronic Load has only one range, say 80 amps and the Current Measurement Accuracy rating in the Spec is 0.1% rdg + 0.1% fs.   Let’s say you are only reading 8 amps.  The accuracy would be as follows:

(8 amps reading x 0.1%) + (80 amps range x 0.1%) = a tolerance of +/- 0.088 Amps.

If the same tolerance is applied to a Load with Multiple Ranges (like the Chroma 63640-80-80, which has 3 ranges, 0.8 amps, 8 amps and 80 amps), and you set the current to 8 amps, you are in the 8 amp range.  The accuracy of the current measurement would be, (8 amps reading x 0.1%) + (8 amps range x 0.1%) = a tolerance of +/- 0.016 amps.

You can clearly see that  having multiple ranges greatly improves the total accuracy of the measurement.  0.088 amps for the single range load versus 0.016 amps for the multi range load.

The Chroma Load Accuracy is 5.5 times greater than the load with a single range.

More Recent Posts

Chroma Unveils Next-Generation Bidirectional DC Power Supplies: 45kW @ 4U High Power Density and 8x Wide-Range Output

In an industry-first, the 62450D-2000HL enables one-key switching between two output ranges within a single unit, effectively providing the output range of two units in one. Users can easily switch between the high voltage (2000V/±60A/±45kW) and high current (650V/±180A/±45kW) range with a single touch, offering up to 8x wider output range capabilities to meet the testing needs of a diverse array of devices under test (DUT).

Read More »

How to Use RF Impedance Measurements to Ensure the Quality of Miniaturized Power Inductors

Most high-frequency and miniaturized power inductors are metal molding-type inductors. The magnetic metal powder surfaces are usually treated with an oxide layer to reduce eddy current losses at high frequencies, enhancing the usable frequency range and reducing losses and temperature rising. However, users typically cannot determine the particle size of the magnetic material and the integrity of the oxide layer after molding and sintering through visual inspection alone.

Read More »