Question: Why is Accuracy and Multiple Ranges important?

Answer:  Most accuracy specifications for Test Equipment are listed as % rdg + % fs.

Example: 0.1% rdg + 0.1% fs

This is a percentage of the reading or measurement + the percentage of the Full Scale for the range you are in.

If a DC Electronic Load has only one range, say 80 amps and the Current Measurement Accuracy rating in the Spec is 0.1% rdg + 0.1% fs.   Let’s say you are only reading 8 amps.  The accuracy would be as follows:

(8 amps reading x 0.1%) + (80 amps range x 0.1%) = a tolerance of +/- 0.088 Amps.

If the same tolerance is applied to a Load with Multiple Ranges (like the Chroma 63640-80-80, which has 3 ranges, 0.8 amps, 8 amps and 80 amps), and you set the current to 8 amps, you are in the 8 amp range.  The accuracy of the current measurement would be, (8 amps reading x 0.1%) + (8 amps range x 0.1%) = a tolerance of +/- 0.016 amps.

You can clearly see that  having multiple ranges greatly improves the total accuracy of the measurement.  0.088 amps for the single range load versus 0.016 amps for the multi range load.

The Chroma Load Accuracy is 5.5 times greater than the load with a single range.

More Recent Posts

Chroma 17050 Offers a Comprehensive BBU Testing Solution

The Chroma 17050 is not only a BBU testing system but also sets a new benchmark for BBU performance- driving the data center industry towards higher efficiency, intelligence, and green energy. Through flexible recipe editing, rapid data acquisition, and real-time monitoring, it helps users comprehensively improve the efficiency and quality of BBU testing.

Read More »

Chroma 1.8MW Bidirectional Power Supply Accelerates the HVDC Era

In AI workloads, instantaneous power surges demand uncompromised reliability, as any brief drop in power can undermine GPU computation. To address this demand, Chroma developed the 62450D-2000HL, a bidirectional DC power supply and regenerative electronic load purpose-engineered for high-voltage, high-power testing.

Read More »

Chroma Implements OCP Standards for BBU Battery Testing to Ensure Reliable Power Backup for AI Servers

Chroma’s regenerative charge/discharge test systems provide comprehensive solutions for both low- and high-voltage BBU modules. In addition to assessing BBU capacity, charge/discharge performance, and long-term durability, Chroma’s 17020C/17050 test systems also enable highly realistic simulation of actual server environments, ensuring accurate and dependable test results.

Read More »